Current Loop Calibrator, Fluke 715 - Fluke
Current loop calibrator The Fluke 715 V/mA Calibrator delivers outstanding performance, durability and reliability. Housed in a rugged Fluke 80 Series DMM housing, the Calibrator is compact, lightweight and easy to carry. splash and has a removable cover for faster battery changes. Brief Description of Functions • Current loop measurement (0-20mA, 4-20mA) with very high accuracy of 0.015% and 1mA resolution • Voltage measurement of process output signals from controllers and converters • Generates or simulates a 24 mA current loop • Generates voltage up to 100mV or 10V • Supplies 24V loop power with simultaneous current measurement...Current loop calibrator
The Fluke 715 V/mA Calibrator delivers outstanding performance, durability and reliability. Housed in a rugged Fluke 80 Series DMM housing, the Calibrator is compact, lightweight and easy to carry. splash and has a removable cover for faster battery changes. Brief Description of Functions
• Current loop measurement (0-20mA, 4-20mA) with very high accuracy of 0.015% and 1mA resolution
• Voltage measurement of process output signals from controllers and converters
• Generates or simulates a 24 mA current loop
• Generates voltage up to 100mV or 10V
• Supplies 24V loop power with simultaneous current measurement
• Improved voltage and current measurement with improved generation accuracy
• Extended battery life
• 250 ohm HART resistor with loop power
• Case with a compartment for storing test leads
• Compatible with the magnetic suspension system included in the accessory kit
• Step and ramp output function
• Customizable zero and span settings
| Functional characteristics | ||
| Current measurement | Range Permission Accuracy | 0 ... 24 mA 0.001 mA 0.02% of reading + 2 DU |
| Current generation | Range Accuracy Power | 0 ... 20 mA or 4 ... 20 mA 0.015% of reading + 2 u 1000 ohm at 24 mA |
| Current simulation | Range Accuracy Requires an external power supply | 0 ... 20 mA or 4 ... 20 mA 0.025% of reading + 2 DU 12 - 30 V |
| Current loop supply | During measurement mA, 24 V | |
| Voltage generation | 0 … 100 mV or 0 … 10 V | |
| Voltage measurement | Range Permission Accuracy | 0 ... 100 mV DC voltage 10 µV / 1 mV 0.02% of reading + 2 DU |
| Environmental requirements | ||
| Working temperature | -10 ºC ... 55 ºC | |
| Storage temperature | -30 ºC ... 60 ºC | |
| Humidity (non-condensing) | 0 … 90% (0 ?C ... 35 ?C); 0 ... 70% (35°C ... 55°C) | |
| Temperature coefficient | -10 ?C ... 18 ?C, 28 ?C ... 55 ?C, ±0.005% of range per ?C | |
| Safety | ||
| Electrical safety standard | CSA C22.2 No. 1010.1:1992 | |
| EMC | EN50082-1:1992 and EN55022:1994 Class B | |
| Max voltage | 30 V | |
| Mechanical and general characteristics | ||
| Size | 152 x 70 x 41mm | |
| Weight | 340 g | |
| Battery type | One 9 V battery type "Krona" | |
Product code 58749184
brand Fluke
State Новое
Manufacturer country США